Abstract
Abstracts
Author Reply
Author's Reply
Book Review
Brief Communication
Case Report
Case Series
Commentary
Continuing Medical Education
Diagnosis
Down the Memory Lane
Editorial
Editorial Board
EDITORIAL BOARD 2026-41-3
Erratum
Faculty
Free papers: Oral Session
Free papers: Poster Session
From Editor's desk
From The Chair, Scientific Committee
Guest Editorial
Image Challenge
In Memoriam
Interesting Image
Interesting Images
Invited Review
Letter to Editor
Letter to the Editor
Letters to Editor
Letters to the Editor
Message
Message by President Elect, SNM, India
Message by President, SNM, India
Messages
Obituary
Oral
ORAL PRESENTATION
Original Article
Pictorial Essay
Pictorial Teaching Essay
POSTER PRESENTATION
President's Message
Presidents’ Wall of Fame
Review
Review Article
Schedule for Paper Presentations
Scientific Program
Secretary's Message
Short Communication
SNM India Guidelines 1.0
SNMI Guidelines
Technical Communication
Technical Note
Generic selectors
Exact matches only
Search in title
Search in content
Post Type Selectors
Search in posts
Search in pages
Filter by Categories
Abstract
Abstracts
Author Reply
Author's Reply
Book Review
Brief Communication
Case Report
Case Series
Commentary
Continuing Medical Education
Diagnosis
Down the Memory Lane
Editorial
Editorial Board
EDITORIAL BOARD 2026-41-3
Erratum
Faculty
Free papers: Oral Session
Free papers: Poster Session
From Editor's desk
From The Chair, Scientific Committee
Guest Editorial
Image Challenge
In Memoriam
Interesting Image
Interesting Images
Invited Review
Letter to Editor
Letter to the Editor
Letters to Editor
Letters to the Editor
Message
Message by President Elect, SNM, India
Message by President, SNM, India
Messages
Obituary
Oral
ORAL PRESENTATION
Original Article
Pictorial Essay
Pictorial Teaching Essay
POSTER PRESENTATION
President's Message
Presidents’ Wall of Fame
Review
Review Article
Schedule for Paper Presentations
Scientific Program
Secretary's Message
Short Communication
SNM India Guidelines 1.0
SNMI Guidelines
Technical Communication
Technical Note
View/Download PDF
Original Article
41 (
3
); 306-310
doi:
10.25259/IJNM_26_2026

Translate:

Validation of Rapid Gel-Clot Testing for the Detection of Bacterial Endotoxin in PET Radiopharmaceuticals with the Standard Cartridge Method- An Economical Solution for Routine Endotoxin Testing

Author affiliations
Department of Biotechnology, Vellore Institute of Technology, Vellore, Tamil Nadu, India.
Department of Neuroimaging and Interventional Radiology, National Institute of Mental Health and Neurosciences, Bengaluru, Karnataka, India.
Article History
Corresponding author(s)

*Corresponding author: Dr. Pardeep Kumar, Department of Neuroimaging and Interventional Radiology, National Institute of Mental Health and Neurosciences, Hosur Rd, Bengaluru, 560029, Karnataka, India. pardeep1182@gmail.com

Article tools · PDF & citations

Abstract

Objectives:

Endotoxin testing is mandatory for radiopharmaceuticals before human administration, as per the World Health Organisation (WHO) guidelines on the use of radiopharmaceuticals in humans. Bacterial endotoxin testing (BET) is mostly performed using either the gel-clot or the chromogenic (cartridge) method.

Material and Methods:

In this study, we have validated the rapid gel clot method with the chromogenic (cartridge) method. We tested three radioactive [18F]FDG samples (n = 3) and three decayed [18F]FDG samples using both a rapid gel assay and a cartridge-based method. Different dilutions were prepared, and the samples were incubated at 37 °C for a pre-set time.

Results:

Both techniques yielded concordant results, and endotoxin was detected below the reference value in the radiopharmaceutical samples. Though the cartridge method was more sensitive and robust, the rapid gel technique requires accurate pipetting and handling. The rapid gel assay is user-friendly and economical for routine testing, especially in developing countries or smallscale facilities.

Conclusion:

The rapid gel assay can serve as an alternative for qualitatively testing endotoxin levels.

Keywords

  • Bacterial endotoxin testing
  • Endotoxins
  • Limulus amoebocyte lysate
  • Lipopolysaccharides
  • Radiopharmaceuticals
Licence
This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-Share Alike 4.0 License, which allows others to remix, transform, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms.

How to cite this article: Sai Teja RB, Muralidhar D, Kumar P. Validation of Rapid Gel-Clot Testing for the Detection of Bacterial Endotoxin in PET Radiopharmaceuticals with the Standard Cartridge Method - An Economical Solution for Routine Endotoxin Testing. Indian J Nucl Med. 2026;41:306-10. doi: 10.25259/IJNM_26_2026

INTRODUCTION

Testing for the presence of bacterial endotoxins in radiopharmaceuticals is crucial to ensuring patient safety before administration during a positron emission tomography (PET) scan. Endotoxins, often called lipopolysaccharides (LPS), are components of the outer cell membrane of the Gram-negative bacteria. They cannot be completely removed by the membrane filters used to purify the PET radiopharmaceutical during synthesis. The higher levels of endotoxins in the radiopharmaceutical can cause fever, chills, hypotension, and might also lead to irreversible shock and death of the patients.[1] Hence, quality control testing of PET radiopharmaceuticals is necessary before their administration to the patients.

The Bacterial Endotoxin Testing (BET) is one such widely used method, approved by the Food and Drug Administration (FDA), for measuring endotoxin levels in pharmaceuticals.[2] BET is based on the principle that the extract from the blood cells (amoebocytes) of the horseshoe crab (Limulus polyphemus), called the Limulus Amoebocyte Lysate (LAL), has the ability to activate a cascade of clotting enzymes and produce an opaque gel in the presence of endotoxins.[3] The Gel-clot method is a traditional, lengthy and cumbersome method of performing BET for radiopharmaceuticals. This method requires manual preparation of the samples, LAL reagent water (LRW), and at least an hour of incubation at 37 °C. This method can yield false-positive and false-negative results due to various factors, including manual error, changes in incubation time, or temperature fluctuations. There are various methods for evaluating endotoxin levels qualitatively and quantitatively. The quantitative method requires additional instruments, such as spectrophotometers, to estimate endotoxin concentration in the formulation.[4]Technological advancements have led to the development of a rapid endotoxin testing system, Endosafe PTSTM (Portable Test System) by Charles River, which features built-in spectrophotometry and provides quantitative results within 15 minutes.[5] An in-built spectrophotometric reader detects endotoxins using a kinetic chromogenic assay in polystyrene LAL cartridge wells. The reagents are loaded and dried in the single-use cartridges; therefore, there is no need to prepare any reagents or standards in this method. The only drawback of this cartridge method, particularly in developing countries like India, is its high cost.

Quality control testing of radiopharmaceuticals is time-sensitive due to the short half-life of the radioisotopes. For instance, Carbon-11 [11C] and Fluorine-18 [18F] have half-lives of approximately 20 and 110 minutes, respectively. Therefore, proper time management is crucial during testing. Hence, we require an affordable and user-friendly method that can be completed within 30 minutes.

Thermo Fisher Scientific’s Pierce™ Rapid Gel-Clot Endotoxin Assay Kits offer a shorter incubation time of 15–35 minutes, which varies by lot and is specified in the Certificate of Analysis. These kits are more cost-effective and time-efficient. Therefore, we have decided to compare this assay with the standard cartridge method.

The primary objective of this study was to compare the performance and feasibility of the gel-clot method (utilising Thermo Fisher Scientific kits) with the cartridge method (Charles River Endosafe PTSTM) and validate the utilisation of kits for routine use.

MATERIAL AND METHODS

18F-FDG sample preparation

The gel clot assay kits [Fig 1] were procured from Thermo Fisher (Pierce™ Rapid Gel Clot Endotoxin Assay Kits). The chromogenic-based assay was performed on Endosafe cartridges, which were read on the EndoSafe NextGen PTS system (Charles River, USA. The pH of the samples was measured either via a pH strip or a pH meter. 18F was produced using an in-house cyclotron (16.5 MeV, GE PETtrace) and [18F]-FDG was synthesised in MX chemistry (GE). Pyrogen-free micro-centrifuge tubes, tips, and endotoxin-free water were obtained from Thermo Fisher, India.

The kit contains three vials: one for the control, one for the sample, and one for the radiopharmaceutical [18F-FDG]sample. FDG: Fluorodeoxyglucose
Fig 1: The kit contains three vials: one for the control, one for the sample, and one for the radiopharmaceutical [18F-FDG]sample. FDG: Fluorodeoxyglucose

Preparation of the sample, kit and endotoxin dilution

The sample [18F]FDG was obtained in a pyrogen-free vial, and the sample was transferred using endotoxin-free tips to the sample tube.

Calculation of maximum valid dilution (MVD)

The formula for the calculation of Endotoxin Limit (EL) is given in equation 1.

EL(EU/mL) = k/M …………………. (Eq. 1)

where, k: Threshold pyrogenic dose =175 EU

M: volume of max recommended at the time of expiration = 7 ml

EL= 175/7 = 25 EU/mL

MVD is the maximum allowable dilution of a sample at which the endotoxin concentration can be determined. The formula for the calculation of MVD is given in equation 2.

MVD=Endotoxin Limit x Product Concentrationλ⋯Eq. 2

where, Endotoxin Limit = 25 EU/mL

Sample Concentration = 1

λ: sensitivity = 0.03 EU/mL

MVD = 25×1/0.03 =833.3 EU/mL

Rapid gel clot testing

The rapid gel kit (Pierce™ Rapid Gel Clot Endotoxin Assay Kits from Thermo Fisher) consisted of the following vials [Fig 1]: a positive product control (PPC), a sample tube (ST), and a radiopharmaceutical sterile vial. Based on the MVD calculation, we prepared four dilutions for testing: 1:10, 1:100, 1:350, and 1:700. We added 500 µL of the sample (18F-FDG) to the sample tube, mixed it, and then transferred 250 µL to the positive control tube, mixing it again. We created a negative control vial by adding 500 µL of water to the sample vial (negative water control, NWC) and then transferring 250 µL to the positive control, which was designated as the positive water control (PWC). After adding the sample (with calculated dilutions from MVD), the tubes were mixed gently by tapping (several times) lightly at the bottom of the tube. The tubes were then gently placed in the incubator (pre-set at 37 °C) for 30 minutes. The incubation time was determined from the certificate of analysis.

The results were interpreted in relation to positive product control: if both PPC and the sample were positive, the endotoxin concentration was considered equal to or greater than the kit's labelled sensitivity.

Cartridge-based testing

In a cartridge-based system [Fig 2], the cartridge consists of four wells: two for the sample and two for the control (duplicate testing). The same dilutions were also tested using the cartridge method. The endotoxin-free water and other consumables were included in the test kit. The cartridge was carefully inserted into the Endosafe machine, and 25.0 µL of the sample was added to each well. After feeding the required values into the endotoxin machine software, the samples were run, and the results appeared on the screen after 25 minutes.

(A) Endosafe NextGen system; (B) Cartridge showing four wells for duplicate sampling of sample (blue arrow) and positive product control (orange arrow).
Fig 2: (A) Endosafe NextGen system; (B) Cartridge showing four wells for duplicate sampling of sample (blue arrow) and positive product control (orange arrow).

Statistics analysis

The batches of radiopharmaceutical were identical and produced on different days; hence, production on different days (n = 3) provided the minimum data needed for basic statistical confidence. The results were assessed qualitatively; therefore, we have used the overall percentage agreement between the two techniques and calculated Cohen’s kappa coefficient.

RESULTS

[18F]-FDG samples were taken from routine production, which was meant for patient use. We tested both radioactive and cold [18F]-FDG samples (in triplicate) to validate the rapid gel-clot assay.

We observed no gel clot during the rapid gel clot assay in the test samples; however, the positive product control showed a gel clot. No clot was observed in the negative water control and negative product control.

Table 1 shows the results of the radioactive [18F]FDG sample reaction tube at an endotoxin concentration of 0.03 EU/ml. The PPC showed gelling, while no gelling was observed in ST after 35 minutes of incubation at 37°C. Therefore, it indicates the absence of endotoxin from the samples. No gelling was observed in the negative control samples as well. A representative image of a gel clot is shown in Fig 3.

Table 1: Results of a hot sample of [18F]FDG (Rapid gel-clot method) on the sample reaction tube at an endotoxin concentration of 0.03 EU/ml.
Dilution Gel vial used Day-1 Day-2 Day-3
1:10 PPC-1 + + +
ST – 1 – – –
1:100 PPC - 2 + + +
ST – 2 – – –
1:350 PPC – 3 + + +
ST – 3 – – –
1:700 PPC -4 + + +
ST – 4 – – –
PWC + + +
NWC – – –

NWC: Negative water control; PWC: Positive water control; PPC: Positive product control; ST: Sample reaction tube, Positive controls and sample reaction tube contained the preparation at a concentration of 0.03EU/ml; PPCs were obtained for all the FDG samples separately at the four different (1:10, 1:100, 1:350 and 1:700) dilutions +: Firm gel formed; −: No gel formed; EU/mL: Endotoxin units per milliliter; FDG: Fluorodeoxyglucose

(A) The positive control (blue capped) vial showed gel clot; (B) The sample vial (red capped) showed no clot after incubation.
Fig 3: (A) The positive control (blue capped) vial showed gel clot; (B) The sample vial (red capped) showed no clot after incubation.

The same samples were tested in a cartridge, and results showed endotoxin levels were less than 1.0 EU/mL after 25 minutes of testing. Table 2 shows the result of [18F]FDG samples with the Cartridge method.

Table 2: Results for [18F]-FDG sample (cartridge method).
Dilution Sample-1 (EU/mL) Sample-2 (EU/mL) Sample-3 (EU/mL)
1:10 < 1.0 < 1.0 < 1.0
1:100 < 1.0 < 1.0 < 1.0
1:350 < 1.0 < 1.0 < 1.0
1:700 < 1.0 < 1.0 < 1.0
WC PWC < 1.0 < 1.0 < 1.0
NWC < 1.0 < 1.0 < 1.0

Cartridge lot number:2667135; NWC: Negative water control; PWC: Positive water control; ; EU/mL: Endotoxin units per milliliter

Another set of tests was conducted using decayed (post-24 hours of production) [18F]FDG samples in the sample reaction tube at an endotoxin concentration of 0.03 EU/mL. Table 3 showed that the PPC (positive product control) was clotted, while no clot was observed in the ST after 35 minutes of incubation at 37°C. This indicated that not all batches contained endotoxin levels greater than 0.03 EU/mL. None of the negative control NWC vials showed a clot. The decayed samples were tested on the cartridge, which showed endotoxin levels below 1.0 EU/mL. There was 100% agreement between the two techniques, and the kappa coefficient was close to 1.0.

Table 3: Results of decayed [18F]FDG (Gel-clot method) on the sample reaction tube at an endotoxin concentration of 0.3 EU/ml.
Dilution Gel vial used Sample-1 Sample-2 Sample-3
1:10 PPC-1 + + +
ST - 1 – – –
1:100 PPC - 2 + + +
ST – 2 – – –
1:350 PPC – 3 + + +
ST – 3 – – –
1:700 PPC -4 + + +
ST – 4 – – –
PWC + + +
NWC – – –

NWC: Negative water control; PWC: Positive water control; PPC: Positive product control; ST: Sample reaction tube; Positive controls and sample reaction tube contained the preparation at a concentration of 0.3EU/ml, PPCs were obtained for all the FDG samples separately at the four different (1:10, 1:100, 1:350 and 1:700) dilutions +: Firm gel formed; −: No gel formed; EU/mL: Endotoxin units per milliliter; FDG: Fluorodeoxyglucose

DISCUSSION

The endotoxin test evaluates the level of fever-producing pyrogens in the formulation. The World Health Organisation (WHO) and international pharmacopoeias mandate the use of the bacterial endotoxin test for radiopharmaceuticals before administering them to humans.[6] Most diagnostic radiopharmaceuticals (positron emitters) are short-lived; therefore, they must be administered within thirty minutes.

The endotoxin cartridge-based method is the most reliable, offering higher sensitivity and faster results than other methods. It is a quantitative method based on chromogenic detection technology. Apart from its advantages, it is an expensive technology and not available in many centres in developing countries. The other method is the gel clot method, which requires multiple endotoxin-free tubes and may take up to 2 hours.[7] Therefore, it is very cumbersome and not feasible in routine work.

In order to develop an affordable and cheap method for routine use. We have validated the rapid gel-clotting assay against standard endotoxin cartridge-based methods. We have done a head-to-head comparison of both techniques and found no difference in results. The maximum volume of 7 ml was selected based on the remaining dose at expiration and the need to dilute for injection. It may vary by site, depending on the concentration remaining at expiry. We have used this concentration based on the daily [18F] FDG remaining at the expiry time. The variable dilution did not impact the outcome or provide any false positives. Therefore, in routine, we can use a 1:700 dilution for testing. The study performed triplicate tests for each batch (n = 3), yielding a total of 9 data points per method. Since both the rapid gel-clot and the cartridge methods yielded identical results across all six radioactive and three decayed samples, the need for a larger sample size is reduced.

Although we have observed that the gel clotting method may yield false positives due to pipetting errors. Therefore, pipetting must be accurate. It is a qualitative method, but it provides insight into the presence of endotoxin in the formulation. There are quantitative methods that utilise chromogenic or turbidimetric techniques, but these methods require additional instruments, such as spectrophotometers, to accurately depict the results. The rapid gel clot assay can be read visually and requires no additional instruments.

When comparing the rapid gel clot assay with the traditional gel clot assay, the latter requires endotoxin-free tubes and takes two hours, whereas the rapid gel clot assay used in our study is user-friendly and takes approximately 35 minutes to complete. Both techniques have their own pros and cons, but rapid gel testing is a cheaper option for endotoxin testing when the Endosafe cartridge method is not available.

CONCLUSION

Rapid gel clot testing is a convenient method for endotoxin testing of radiopharmaceuticals, particularly for facilities without an Endosafe cartridge setup. The rapid gel testing is user-friendly and takes a maximum of 35 minutes. Our study has validated the rapid gel clot method with the standard Endosafe cartridge method.

Author contributions:

ST: Experimentation, drafting manuscript; DM: Experimentation, supervision and writing manuscript; PK: Designed the study and final manuscript preparation

Ethical approval:

Institutional Review Board approval is not required as it did not involve any patient data or fluid sample from patients.

Declaration of patient consent:

Patient's consent not required as there are no patients in this study.

Conflicts of interest:

There are no conflicts of interest.

Use of artificial intelligence (AI)-assisted technology for manuscript preparation:

The authors confirm that there was no use of artificial intelligence (AI)-assisted technology for assisting in the writing or editing of the manuscript and no images were manipulated using AI.

Financial support and sponsorship: Nil.

References

  1. , , , , , , et al. Endotoxin-induced physiological and psychological sickness responses in healthy humans: Insights into the post-acute phase. Neuroimmunomodulation. 2023;30:268-76.
    [CrossRef] [PubMed] [Google Scholar]
  2. . IAEA/WHO guideline on good manufacturing practices for investigational radiopharmaceutical products. WHO expert committee on specifications for pharmaceutical preparations. Annex WHO Tech Rep Ser. 2022;1044:171-94.
    [Google Scholar]
  3. , . The role of endotoxin in the extracellular coagulation of limulus blood. Bull Johns Hopkins Hosp. 1964;115:265-74.
    [Google Scholar]
  4. , , . Quantitative comparison of in vitro and in vivo methods for the detection of endotoxin. J Lab Clin Med. 1971;78:138-48.
    [Google Scholar]
  5. . Endosafe PTS: A portable, kinetic chromogenic LAL test system .
    [Google Scholar]
  6. , , , , , , et al. Guideline on current good radiopharmacy practice (cGRPP) for the small-scale preparation of radiopharmaceuticals. EJNMMI Radiopharm Chem. 2021;6:8.
    [CrossRef] [PubMed] [Google Scholar]
  7. , , , , , , et al. A multicenter comparison study between the Endosafe PTS rapid-release testing system and traditional methods for detecting endotoxin in cell-therapy products. Cytotherapy. 2008;10:427-35.
    [CrossRef] [PubMed] [Google Scholar]
Show Sections