Translate this page into:
Comparison of Image Quality of Different Radionuclides Technetium-99m, Samarium-153, and Iodine-123
Address for correspondence: Dr. Youssef Bouzekraoui, Department of Physics, LPHE, Modeling and Simulations, Faculty of Science, Mohammed V University, Rabat, Morocco. E-mail: youssef0fsr@gmail.com
This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
This article was originally published by Medknow Publications & Media Pvt Ltd and was migrated to Scientific Scholar after the change of Publisher.
Abstract
Introduction:
The choice of the radionuclide has a key role in nuclear medicine which appearing the lowest scatter fraction. In addition, the presence of penetrated and scattered photons from collimator in single-photon emission computed tomography images degrades resolution and contrast. Thus, image quality depends on sensitivity and resolution of the collimator–detector system. The goal of this study was to compare the image quality that can be achieved by three radionuclides: technetium-99 m (Tc-99 m), iodine-123 (I-123), and samarium-153 (Sm-153).
Materials and Methods:
Tc-99 m and Sm-153 were imaged with low-energy high resolution (LEHR) collimator, while I-123 was imaged with medium-energy (ME) collimator. We modeled the Siemens Symbia Medical system using Monte Carlo simulation code SIMIND. The imaging characteristics of each radionuclide were investigated by simulated data: point spread function, sensitivity (Cps/MBq) and geometric, penetration and scattering distribution.
Results:
Tc-99 m and Sm-153 give best and results with LEHR collimator for spatial resolution (full width at half maximum [FWHM] = 3.19 mm; full width at tenth maximum [FWTM] = 6.73 mm) and (FWHM = 3.22 mm; FWTM = 7.39 mm), respectively. Whereas, I-123 provided with ME collimator a lower resolution (FWHM = 4.89 mm; FWTM = 9.89 mm). The sensitivity recorded by Tc-99 m, Sm-153, and I-153 were (31.21 Cps/MBq), (10.16 Cps/MBq), and (51.22 Cps/MBq), respectively.
Conclusion:
Tc-99 m and Sm-153 give the best and generally similar imaging properties with LEHR. For I-123, the ME collimator helps lowering the influence of high-energy gamma rays.
Keywords
Geometric
imaging
penetration
resolution
scattering
sensitivity
SIMIND
Introduction
In nuclear medicine, the radionuclides technetium-99 m (Tc-99 m) and iodine-123 (I-123) are most commonly used in single-photon emission computed tomography imaging. While, Samarium-153 (Sm-153) emits gamma radiation at 103 keV and has a physical half-life of 46.7 h, which is well suited for scintigraphic imaging.[1] Whereas, Sm-153 possesses both therapeutic beta and diagnostic gamma radiations, making it an alternative radionuclide to Y-90 in liver cancer treatment.[2] The Sm-153 imaging was useful in gastrointestinal scintigraphy.[3] In addition, Sm-153 has been used in various scintigraphic studies.[45] Sm-153 has average, and maximum beta particle ranges in water of 0.5 mm and 3.0 mm, respectively. It is undeniable that the image quality strongly depends on the types of detected photons in the photopeak window. This photon retching the detector were classified as: the geometric component (passed without any interaction inside the collimator), the penetration component (passed through septa without attenuation), or the scatter component (scattered in the septa). In addition, photons that are absorbed in the collimator septa produce an X-ray component by the photoelectric effect.[6] Only the first component provides the correct information. The penetration and scattering effects in collimator degrade contrast, resolution, and quantification. As well as, image quality and quantification accuracy are affected by these factors.[7] Gamma camera cannot classify the detected photons into geometric, penetrated, or scattered photons. To solve this difficulty, we used Monte Carlo simulation technique to assess the geometric, penetration, and scatter contribution inside the photopeak window.[8] Whereas there are few studies related with the assessment of geometric, penetration, and scatter components.[9101112]
The aim of this study was to evaluate the image quality for each radionuclide. We model the Siemens Symbia medical system Gamma Camera, using Monte Carlo SIMIND code.[1314151617] Moreover, evaluate the geometric, scattered, and septal penetration components. We also evaluate the resolution and sensitivity (Cps/MBq).
Materials and Methods
Images were acquired on a Siemens Medical System Symbia using both a medium-energy (ME), low-energy high resolution (LEHR), and ME collimators [Table 1]. I-123 was imaged with ME collimator, whereas Tc-99 m and Sm-153 were imaged with LEHR [Table 2]. Therefore, a point source has been placed at 12 cm from the detector surface and located in horizontal cylindrical water phantom of dimension 16 cm × 22 cm × 22 cm. We have modeled the detector with a radius equal to 25 cm having 2.54 cm NaI (Tl) crystal thickness. We made the intrinsic spatial resolution of 0.34 cm and energy resolution of 8.80% at 140 keV. We have used, the SIMIND Monte Carlo code developed by Michael Ljungberg.
| Low energy | ME | LEHR |
|---|---|---|
| Geometric of hole | Hexagonal | Hexagonal |
| Length of hole (mm) | 4.064 | 2.405 |
| Septal thickness (mm) | 0.114 | 0.016 |
| Diameter of hole (mm) | 0.294 | 0.111 |
ME: Medium-energy, LEHR: Low-energy high resolution
| Tc-99m | Sm-153 | I-123 | |||
|---|---|---|---|---|---|
| Energy (keV) | Abundance (%) | Energy (keV) | Abundance (%) | Energy (keV) | Abundance (%) |
| 20.67 | 1.12 | 40.9 | 16.6 | 27.2 | 24.69 |
| 21.02 | 0.18 | 41.54 | 30 | 27.47 | 45.98 |
| 140.51 | 88.5 | 47.11 | 9.45 | 31.1 | 13.16 |
| - | - | 48.38 | 2.44 | 31.76 | 2.86 |
| - | - | 69.67 | 4.69 | −158.97 | 83.25 |
| - | - | 75.42 | 0.17 | 346.35 | 0.13 |
| - | - | 83.37 | 0.19 | 440.02 | 0.42 |
| - | - | 89.49 | 0.16 | 505.33 | 0.27 |
| - | - | 97.43 | 0.77 | 528.96 | 1.28 |
| - | - | 103.18 | 29.19 | 538.54 | 0.38 |
Tc-99m: Technetium-99m, Sm-153: Samarium-153, I-123: Iodine-123
Images were acquired using a 128 × 128 matrix and a pixel size = 0.34 mm × 0.34 mm. The images created by SIMIND in ImageJ software, National Institutes of Health, and the Laboratory for Optical and Computational Instrumentation (University of Wisconsin).[18]
Table 2 shows photons energies and intensities of I-123, Tc-99 m, and Sm-153 radionuclides decay. For I-123 and Sm-153, ten peaks have been simulated, and three peaks have also been simulated for Tc-99 m [Table 2]. Simulations were carried out for two separate conditions: in air and with scattering medium (cylinder of water). In this simulation study, the main energy window was centered on the main energy peak of radionuclides for width of 20% of this energy (I-123 [159 keV]: 143–175 keV, Tc-99 m [141 keV]: 131–151 keV, and Sm-153 [103 keV]: 93–113 keV). The point spread function (PSF) was studied for each radionuclide. The full width at half maximum (FWHM) and the full width at tenth maximum (FWTM) were computed from the PSF to quantify the resolution. All imaging parameters have been compared with or without a scattering medium.
Results and Discussion
The detected spectrum of radionuclides without collimator is shown in Figure 1. The energies are shown the energies of the main emission peaks of the isotopes.

The different fractions detected within 20% peak energy window for all radionuclides are shown in Table 3. In water, the geometric fraction is large for Tc-99 m compared to the two other radionuclides. Whereas, in air this fraction, it is approximately the same for the three radionuclides. Because, these components depend on the energy of photons, object under study and collimator design parameters.
| Phantom | Isotopes | Collimators | Geometric (%) | Penetration | Scatter (%) |
|---|---|---|---|---|---|
| Water | Tc-99m | LEHR | 90.92 | 5.99 | 3.09 |
| Sm-153 | LEHR | 85 | 10.46 | 4.36 | |
| I-123 | ME | 88.08 | 7.4 | 4.52 | |
| Air | Tc-99m | LEHR | 90.09 | 6.61 | 3.3 |
| Sm-153 | LEHR | 89.49 | 7.13 | 3.24 | |
| I-123 | ME | 90.43 | 6.03 | 3.54 |
ME: Medium-energy, LEHR: Low-energy high resolution, Tc-99m: Technetium-99m, Sm-153: Samarium-153, I-123: Iodine-123
Figure 2 shows images of a point source for different radionuclides in water and air. The fogginess in the images formed by I-123 in water and air caused by scatter and penetration components due to high-energy gamma ray's emissions. The images formed by Sm-153 and TC-99 m show clearly a good quality.

Resolution and sensitivity (Cps/MBq) study
The PSF obtained for all of the isotopes, in air and water, are shown in Figure 3. Tc-99 m and Sm-153 offer a better resolution than the I-123 in both cases. The curve of I-123 with the ME collimator shows the effects of the septa which lowers the resolution.

Images quality depends on resolution and also by sensitivity of the detector collimator system. Figure 4 shows the comparison of sensitivity for three radionuclides.

Figure 4 shows that for each isotope the water phantom decreases the sensitivity. For I-123, the ME produced the high sensitivity compared with LEHR for two other radionuclides.
Quantification of the resolution
FWHM and FWTM were calculated from profiles through the source images.
FWHM and FWTM are shown in Table 4; it shows that, for each isotope, the water phantom increases FWTM, while FWHM remains roughly the same. FWHM was better for the LEHR then for the ME collimator. In addition, FWHM was similar for both Tc-99 m and Sm-153.
| Phantom | Isotopes | FWHM (mm) | FWTM (mm) | Sensitivity (cps/MBq) |
|---|---|---|---|---|
| Water | Tc-99m | 3.19 | 6.73 | 31.21 |
| Sm-153 | 3.22 | 7.39 | 10.16 | |
| I-123 | 4.89 | 9.89 | 51.22 | |
| Air | Tc-99m | 3.01 | 5.41 | 104.9 |
| Sm-153 | 3 | 5.39 | 34.48 | |
| I-123 | 4.62 | 7.78 | 162.7 |
FWHM: Full width at half maximum, FWTM: Full width at tenth maximum, Tc-99m: Technetium-99m, Sm-153: Samarium-153, I-123: Iodine-123
Conclusion
The effect of penetrated and scattered photons has been studied both qualitatively and quantitatively for Sm-153 and Tc-99 m and I-123. Since the results of the simulation show that the Sm-153 and Tc-99 m produced, generally, the best and similar imaging characteristics with LEHR collimator. However, the sensitivity obtained, with air and water phantom, was more than that of the other one. For I-123 imaging, the use of an ME collimator is important when the resolution is not required.
Financial support and sponsorship
Nil.
Conflicts of interest
There are no conflicts of interest.
References
- High specific activity [samarium-153] EDTA for imaging of experimental tumor models. J Nucl Med. 1989;30:202-8.
- [Google Scholar]
- Neutron activated samarium-153 microparticles for transarterial radioembolization of liver tumour with post-procedure imaging capabilities. PLoS One. 2015;10:e0138106.
- [Google Scholar]
- Production and first use of 153SmCl3-ion exchange resin capsule formulation for assessing gastrointestinal motility. Appl Radiat Isot. 2012;70:450-5.
- [Google Scholar]
- Neutron activation-based gamma scintigraphy in pharmacoscintigraphic evaluation of an egalet constant-release drug delivery system. Int J Pharm. 2004;281:3-10.
- [Google Scholar]
- In vivo evaluation of enteric-coated naproxen tablets using gamma scintigraphy. Pharm Res. 1992;9:1436-41.
- [Google Scholar]
- Approximation of approximation of hexagonal holes by square holes in Monte Carlo simulation of gamma-camera collimation. IEEE Trans Nucl Sci. 2002;49:2186-95.
- [Google Scholar]
- Monte Carlo assessment of geometric, scatter and septal penetration components in DST-XLi HEGP collimator. IFMBE Proc. 2008;22:2479-82.
- [Google Scholar]
- A new collimator simulation in SIMIND based on the delta scattering technique. IEEE Trans Nucl Sci. 2005;52:1370-5.
- [Google Scholar]
- Evaluation of the geometric, scatter, and septal penetration components in fan-beam collimators using Monte Carlo simulation. IEEE Trans Nucl Sci. 2002;49:12-6.
- [Google Scholar]
- Comparison of image quality of different iodine isotopes (I-123, I-124 and I-131) Cancer Biother Radiopharm. 2007;22:423-30.
- [Google Scholar]
- Monte Carlo modeling of gamma cameras for I-131 imaging in targeted radiotherapy. Cancer Biother Radiopharm. 2005;20:77-84.
- [Google Scholar]
- Characterization of parallel-hole collimator using Monte Carlo simulation. Indian J Nucl Med. 2015;30:128-34.
- [Google Scholar]
- The SIMIND Monte Carlo Program Home Page. Available from: https://www.msf.lu.se/forskning/the-simindmonte-carlo-program
- [Google Scholar]
- Development and evaluation of an improved quantitative (90) Y bremsstrahlung SPECT method. Med Phys. 2012;39:2346-58.
- [Google Scholar]
- A collimator optimization method for quantitative imaging: Application to Y-90 bremsstrahlung SPECT. Med Phys. 2013;40:082504.
- [Google Scholar]
- SIMIND monte carlo simulation of a single photon emission CT. J Med Phys. 2010;35:42-7.
- [Google Scholar]
- Collimator and energy window optimization for 90Y bremsstrahlung SPECT imaging: A SIMIND monte carlo study. Appl Radiat Isot. 2016;108:124-8.
- [Google Scholar]
